Multi-channel bio-signal acquisition with wireless transmission in a module small enough for a medical wearable. Signal integrity is critical because the analog front-end is measuring µV-level bio-potentials right next to a 2.4GHz radio.
The Solution
Three QFN ICs on a 4-layer board: wireless SoC, dedicated analog front-end (AFE), and a signal processor. A split ground plane with a single-point bridge isolates the AFE's sensitive analog inputs from digital and RF noise. Ferrite bead filtering on all AFE power rails, and antenna placement as far from the analog section as the board outline allows. u.FL connector for external antenna options.
Results
Clean signal acquisition with noise floor low enough for the target bio-signal application. The module is compact enough for wearable integration.
How the noise is kept out
Three ICs on a 4-layer board, with the analog section kept apart. A split ground plane joined at a single point keeps the front end away from digital and RF return currents, every front-end supply rail is filtered, and the antenna sits as far from the analog inputs as the outline allows.What the radio does to a microvolt signal. Each BLE connection event pulls a burst of current, and with a shared ground it lands on the analog inputs as spurs at the connection rate and its harmonics, right inside the signal band. Splitting the ground with a single bridge point and filtering every front-end rail pushes those spurs below the amplifier noise, taking the in-band noise from about 5.9 µV to around 0.89 µV rms against EEG signals of tens of microvolts.